Cirrus-logic CDB5484U Bedienungsanleitung Seite 34

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 47
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 33
CDB5484U
34 DS919DB5
Figure 27. Histogram Analysis
2.8.1.12 Frequency Domain Information
The following describes the indicators associated with FFT (Fast-Fourier Transform) analysis. FFT data
can be plotted in the Data Collection window by setting the analysis type selector to FFT. See Figure 28.
The FFT information includes:
FREQUENCY: Displays the x-axis value of the cursor on the FFT display.
MAGNITUDE: Displays the y-axis value of the cursor on the FFT display.
S/PN: Indicates the signal-to-peak noise ratio (decibels).
SINAD: Indicates the signal-plus-noise-plus-distortion to noise-plus-distortion ratio (decibels).
S/D: Indicates the signal-to-distortion ratio, 4 harmonics are used in the calculations (decibels).
SNR: Indicates the signal-to-noise ratio, first 4 harmonics are not included (decibels).
FS-PdB: Indicates the full-scale to signal Ratio (decibels).
Figure 28. FFT Analysis
Seitenansicht 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 46 47